Conmed Aspen Excalibur Specifications | Mode | Max. Power | Max. P-P Voltage | Crest Factor | Load (ohms) | | Pure cut | 300 W | 2000 | 1.8 | 300 | | Blend | 180 W | 2000 | 2.5 | 300 | | Coag Std | 20 W | 6500 | 6.0 | 300 | | Coag Spray | 80 W | 10500 | 9.0 | 500 | | Bipolar Coag | 50 W | 400 | 2.1 | 50 | | Bipolar Cut | 50 W | 400 | 1.6 | 50 | Operating Frequency: Cut: 416.7 kHz ±0.01%. Coag: 540 kHz ±10%. Bipolar: 1.05 MHz ±0.01% Pulse Repetition Frequency: Blend, Spray and Bipolar Coag: 20 kHz ± 0.1%. Standard Coag: 20 kHz ±0.1% Input Voltage: 104-127 VRMS. 60 Hz calibrated at 120 Input Current: Idle: 0.8 A. Pure Cut at 300: 6.5 A. Coag at 120: 2.8 A. Line Frequency Leakage: Patient connected to neutral < 10 µA. Chassis to neutral < 25 µA RF Resolution: 99% Minimum Power display Accuracy: Within 10%, or 3 watts at rated load. Weight: 40 lbs. (18.2 kg) Dimensions: 6.25”H x 15.9”W x 22.75”D (15.9 x 40.4 x 57.8 cm) Click the 'Get Phone Number" button above and call us for more information.
|